Volume discount: 5 pcs 2% · 10 pcs 3% · 25 pcs 5%The fridge failed on Friday evening and no one noticed until Monday morningLegionella control that doesn’t end up in a binder: timestamped temperatures straight from the tapMoisture that isn’t visible on the surface becomes most expensive when it’s discovered after the floor is laidA Surface That Looks Clean Can Still Be Dirty: How ATP Measurement Shows the Result in Ten SecondsStop guessing the salt in your brine: read the concentration in a few secondsThree known temperature points are enough to tell whether your thermometer still reads correctlyHow to Know Your Thermometer Still Reads Correctly: Checking Type K Instruments at Three Known PointsOne drop is enough to see whether the coolant concentration is correct, without interpreting a borderline by eyeA spot check doesn’t tell you what happened in the cold room overnight between two inspectionsHow to Measure Small, Hot Surfaces from a Safe Distance Without GuessingVolume discount: 5 pcs 2% · 10 pcs 3% · 25 pcs 5%The fridge failed on Friday evening and no one noticed until Monday morningLegionella control that doesn’t end up in a binder: timestamped temperatures straight from the tapMoisture that isn’t visible on the surface becomes most expensive when it’s discovered after the floor is laidA Surface That Looks Clean Can Still Be Dirty: How ATP Measurement Shows the Result in Ten SecondsStop guessing the salt in your brine: read the concentration in a few secondsThree known temperature points are enough to tell whether your thermometer still reads correctlyHow to Know Your Thermometer Still Reads Correctly: Checking Type K Instruments at Three Known PointsOne drop is enough to see whether the coolant concentration is correct, without interpreting a borderline by eyeA spot check doesn’t tell you what happened in the cold room overnight between two inspectionsHow to Measure Small, Hot Surfaces from a Safe Distance Without Guessing

When the thermometer reads wrong, the question isn’t how much, but whether it’s the instrument or the sensor

Published 23 August 2026 at 17:41

A thermometer in your food safety checks reads 2.4 degrees too low. That’s not a disaster in itself — but the question that decides what to do about it rarely gets asked: is the fault in the instrument or in the probe? Replace the probe and the fault is still there, and you’ve wasted money. Send the instrument off for calibration when the fault is actually a damaged probe cable, and it comes back passed — with the next measurement just as wrong as the last.

ETI MicroCal 1 calibrated temperature simulator for thermocouple types K J T R N S and E

What it is

A calibrated temperature simulator. Instead of heating a probe, it feeds in exactly the signal a probe would give at a chosen temperature — so what you’re measuring is the instrument, and nothing else.

  • Simulates thermocouple types K, J, T, R, N, S and E, with no conversion tables needed
  • 12 adjustable temperature points
  • Selectable cold junction compensation (CJC), internal or external
  • Range per type, e.g. K: -200 to 1372 degrees C, T: -270 to 400, R and S: 0 to 1768
  • All types conform to BS EN 60584-2
  • 10 mm LCD with alphanumeric row, choice of degrees C or F, auto shut-off
  • Supplied with type K cable and a five-point UKAS calibration certificate

The problem it solves

A simulator splits the measurement chain into two parts that can be troubleshot separately. Connect the simulator in place of the probe, and if the instrument reads correctly, the instrument is sound and the fault lies in the probe or cable. If it reads wrong, the fault is in the instrument. It’s the same logic as swapping a suspect component for a known-good one — and it’s the only check that gives an unambiguous answer, instead of one combined error for the whole chain.

This is not the same as a dry-block calibrator, and the two don’t replace each other. A dry block heats an actual probe to an actual temperature and so checks the whole chain as it’s really used — including the probe’s own errors. A simulator skips the probe entirely. One answers is my measurement correct, the other answers where is the fault.

Cold junction compensation is the underrated detail. A thermocouple doesn’t give an absolute temperature, only a voltage corresponding to the difference between the measuring point and the connection at the instrument. That means the instrument has to know its own connection temperature and subtract it out. When that compensation goes wrong, the thermometer reads differently depending on how warm the room is — an error that’s particularly deceptive because it disappears as soon as someone checks the instrument in a different room. With selectable internal or external CJC, that function can be tested rather than simply assumed.

Seven probe types in one instrument also means operations with mixed equipment don’t need one tool per type — and never have to look up millivolt tables by hand, which is one of the places checks otherwise go wrong.

Three typical use cases

  • Food safety checks: regular verification that kitchen and storage thermometers read correctly, with documentation that holds up at an audit.
  • Maintenance and service: field troubleshooting, where you need to know whether it’s the meter or the probe that needs replacing before ordering a spare part.
  • Internal calibration work: as a reference for periodic checks of multiple instruments, between external calibrations.

Why it pays off

External calibration of every thermometer, every year, is both a cost and a period when the instrument is out of service. With a calibrated simulator on hand, frequent checks become a routine that takes minutes, and external calibration only needs to cover the simulator — the reference — instead of every individual meter.

The bigger saving, though, is the one that never shows up on an invoice. A thermometer that has drifted out of spec gets caught at the next routine check instead of at an audit, and then the question concerns a single meter — not whether every temperature logged since the last calibration can still be trusted.

Read more about the ETI MicroCal 1 temperature simulator →