Allergen-safe digital pocket thermometer, measuring range -50°C to +200°C
366 SEK
The KM400AP waterproof allergy thermometer is color-coded for food allergen procedures and features a thin tip for fast response times.
Volume discount: 5 pcs: 2% · 10 pcs: 3% · 25 pcs: 5%
- Free shipping on orders over 3000 SEK excl. VAT
- Measuring instruments and hygiene tests for self-monitoring since 2003
- Call us +46 8 753 00 04
Category: Digital pocket thermometers
Other models in the series 9
Description
The KM400AP waterproof allergen thermometer is color-coded for food allergen procedures and features a thin probe for fast response times.
The pen-style KM400AP Waterproof Allergen Thermometer is purple, the agreed-upon color in the food service industry for kitchens to identify allergen-free foods.
If your kitchen regularly handles milk, eggs, fish, peanuts, tree nuts (walnuts, almonds), wheat, soybeans, and shellfish (shrimp, crab) or other food allergens, preventing cross-contamination becomes increasingly important. HACCP plans now include detailed regulations regarding the segregation of foods or ingredients that may cause allergies and the use of separate utensils for different foods.
The KM400AP waterproof allergen thermometer features a 1.5 mm thin probe for a fast response, along with a hold function on the thermometer that allows the user to temporarily fix the reading on the screen while noting a measurement.
Features:
- Automatic shut-off
- Data hold
- BioCote antimicrobial protection
- NSF listed
- Color-coded for food safety
- Purple color is for allergens
- Thin probe for fast response
- Easy-to-read LCD display
- On/off button
- 2.75 inch (75 mm) stainless steel shaft
- Storage case with clip
- 1.5 volt battery included
User Manual
Press the ”ON/OFF” button to start or stop the operation of the KM400AP. The unit automatically turns off after 35 minutes to conserve battery life. A low battery is indicated by ”LOB” flashing alternately with a reading. Replace the (AB13) (1.5V LR44) battery immediately.
Switching between °C and °F
With the unit turned off, press and hold the ”ON/OFF” button, and the display will switch between ”°F” and ”°C”. Release the ”ON/OFF‘ button when the desired mode is displayed.
Data Hold Function
To hold a displayed reading, press the ”D-H” (Data Hold) button. The reading flashes when Data Hold is active. Press the ”D-H” (Data Hold) button again to return to normal operation.
Battery Replacement
Use a coin to open the compartment. Replace with a (AB13) (1.5V LR44) battery and close the cover.
Field Calibration Procedures
1. The KM400AP can only be calibrated in an ice slurry (60% crushed ice and 40% tap water) or a water bath. Do not attempt to calibrate by any other method.
2. Immerse the thermometer stem to at least 1/2 of its length in the ice slurry.
3. Wait until the reading stabilizes (typically ±1° of 0 °C / ±2° of +32° F).
4. Press and hold the D-H button for 8 seconds to initiate calibration.
5. Calibration is displayed for two seconds, and calibration at 0 °C / +32 °F is complete.
6. Remove the thermometer from the ice water and continue measuring.
Note: CAL is displayed when the D-H button is pressed for 8 seconds, but the thermometer will not calibrate unless the probe is immersed in ice water.