Checklite ATP Eliminating Kit for 250 Analyses
3.818 SEK
- Specially designed ATP-degrading enzymes
- Highly sensitive microbial biomass analysis
- Easy to use
- Fast performance
Volume discount: 5 pcs: 2% · 10 pcs: 3% · 25 pcs: 5%
- Free shipping on orders over 3000 SEK excl. VAT
- Measuring instruments and hygiene tests for self-monitoring since 2003
- Call us +46 8 753 00 04
Description
78-1002660-001 Test hygiene > Checklite
The CheckLite ATP Eliminating Kit is used for microbial biomass studies based on the ATP measurement method. All living cells, including microorganisms, use ATP as an energy source. Therefore, the total cell mass can be determined by measuring bioluminescence using the luciferase reaction. This is normally done by extracting ATP from the sample using an ATP-releasing reagent, such as in the CheckLite 250 Plus. However, this requires that extracellular ATP levels are low.
A major issue is that extracellular ATP can be several orders of magnitude higher in certain samples, which hinders the sensitive detection of intracellular ATP. The ATP-eliminating reagent reduces extracellular ATP to very low concentrations, enabling targeted detection of bacteria.
1 x 250 analyses
Test duration: 30-40 minutes
Storage: 2-8 °C, DO NOT FREEZE
Kit Composition
ATP-eliminating reagent:
5 vials marked red-pink
Reconstitution buffer for ATP-eliminating reagent:
5 vials marked yellow
Reagent Preparation
The ATP-eliminating reagent is stored under vacuum in a red-pink marked vial.
Pour the reconstitution buffer from the yellow-marked vial into the opened red-pink marked vial and let it stand at room temperature for a few minutes.
Gently swirl the vial to avoid foam formation until the contents are completely dissolved.
Do not touch the rim of the vial or the top of the rubber stopper directly with your hands, as this may sometimes increase the reagent’s blank value.
One vial of ATP-eliminating reagent can be used for more than 50 analyses under normal conditions.
Notes on Sample Handling
The concentration of ATP in living cells changes rapidly. Therefore, extract ATP from the cells or freeze the cells immediately after sampling. Even during extraction, it may be necessary to inhibit or destroy ATP-degrading enzymes to prevent the enzymes from reducing the ATP concentration.
Sample Pre-treatment
3.1. Solid samples:
Process the sample with a stomacher or a homogenizer and then use the supernatant for the subsequent measurement.
3.2. Liquid samples:
If the sample solution is turbid, colored, or contains inhibiting ions, such as Cl-, dilute the sample as needed.
ATP degradation of samples
Place 1 ml of sample in the test tube.
Add 0.1 ml of ATP-elimination reagent to the test tube and mix well.
Let the test tube stand at room temperature for 30 to 40 minutes. It is important to keep the reaction time constant to achieve good reproducibility. If you are testing multiple samples, each test tube must stand at room temperature for the same duration.
After the room temperature reaction described above, the samples are used for the bioluminescence analysis below.