Calibration Kit with LED Control Set for Lumitester PD20 / PD30 / SMART, 902750000 Japan
8.966 SEK
- Electric method with LED
- No radioactive fluorescence
- Stable light generated by automatic light control
- Rechargeable battery
Volume discount: 5 pcs: 2% · 10 pcs: 3% · 25 pcs: 5%
- Free shipping on orders over 3000 SEK excl. VAT
- Measuring instruments and hygiene tests for self-monitoring since 2003
- Call us +46 8 753 00 04
Description
Guarantee your quality control
The Lumitester Control Kit makes your hygiene checks even more secure. It verifies the accuracy of your Lumitester in just 30 seconds. It enables continuous self-monitoring and verifies both the perfect condition and correct function of the Luminometer.
The control kit consists of a control stick (positive control), a negative control, and a charger with a power supply unit. It can only be used with Lumitesters from models PD20, PD30, and Smart.
You can use the control stick to validate whether the Lumitester’s measurement results still fall within specification. Unlike competitors’ products, no radioactive substances are used.
The positive control is based on an artificial green LED light source that emits an exactly defined, pre-set, and calibrated amount of light. This amount of light is measured by the Lumitester in the same way as with the LucipacPen stick. If the target value between 80 and 120% printed on the control kit is restored, the Lumitester is in perfect condition. In case of a higher deviation, the Lumitester’s chamber must be cleaned according to the procedure described in the user manual.
1 control kit
Test duration: 10 sec
Kit composition
1 Light source (positive stick)
1 Empty tube (negative stick, typical 0 RLU)
1 AC/DC adapter for charging (output DC 5V 3A)
1 Power cord, Type A plug
1 Power cord, Type C plug
1 Bag
1. Connect the control plug to the power supply and charge for at least 30 seconds before use
2. Turn on the Lumitester and insert the control stick into the measuring chamber of the Lumitester.
3. Close the device. Press the ”Enter” button and read the result from the display. To check the measurement results, you can find the target value in RLU (Relative Light Units) on the control stick’s label. If an RLU value between 80 and 120% of the printed target value is achieved, and a value under 10 RLU is measured in the negative control, the Lumitester provides perfect results. If this is not the case, follow the troubleshooting in the manual.
Specification
Guarantee your quality control
The Lumitester Control Kit makes your hygiene checks even more secure. It verifies the accuracy of your Lumitester in just 30 seconds. It enables continuous self-monitoring and verifies both the perfect condition and correct function of the Luminometer.
The control kit consists of a control stick (positive control), a negative control, and a charger with a power supply unit. It can only be used with Lumitesters from models PD20, PD30, and Smart.
You can use the control stick to validate whether the Lumitester’s measurement results still fall within specification. Unlike competitors’ products, no radioactive substances are used.
The positive control is based on an artificial green LED light source that emits an exactly defined, pre-set, and calibrated amount of light. This amount of light is measured by the Lumitester in the same way as with the LucipacPen stick. If the target value between 80 and 120% printed on the control kit is restored, the Lumitester is in perfect condition. In case of a higher deviation, the Lumitester’s chamber must be cleaned according to the procedure described in the user manual.
1 control kit
Test duration: 10 sec
Kit composition
1 Light source (positive stick)
1 Empty tube (negative stick, typical 0 RLU)
1 AC/DC adapter for charging (output DC 5V 3A)
1 Power cord, Type A plug
1 Power cord, Type C plug
1 Bag
1. Connect the control plug to the power supply and charge for at least 30 seconds before use
2. Turn on the Lumitester and insert the control stick into the measuring chamber of the Lumitester.
3. Close the device. Press the ”Enter” button and read the result from the display. To check the measurement results, you can find the target value in RLU (Relative Light Units) on the control stick’s label. If an RLU value between 80 and 120% of the printed target value is achieved, and a value under 10 RLU is measured in the negative control, the Lumitester provides perfect results. If this is not the case, follow the troubleshooting in the manual.